FEATURES:
Meets ISO 7027/EPA 180.1 for Turbidity Meters
Low Level Detection of Turbidity, Chlorine, and Color
Programmable Auto-Shut Off
Log up to 4000 time and date stamped recordings
Supports Six Languages
Portable Chlorine Meters Product Description
The TC-3000 Portable Chlorine Meter is ideally suited for both low-level drinking water applications as well as monitoring high turbidity in the field. The portable chlorine meters have a six-detector design with special focusing optics allowing for long-term stability over a wide range of operational conditions. The portable chlorine meters use the unique detector arrangement to measure Chlorine, Turbidity, and Color. The portable chlorine meters come in two designs: one with a tungsten lamp to meet EPA 180.1 turbidity specifications, the other comes with a IR LED emitter and meets the ISO 7027 turbidity standard. The portable chlorine meters use a tube-positioning ring to limit tube variability enabling maximum sensitivity and accuracy.
The TC-3000 Portable Chlorine Meter uses an advanced micro controller which allows for advanced calibration algorithms and support of up 6 languages including English, French, Spanish, Japanese, Portuguese, and Italian. The portable chlorine meters can record up to 4000 time and date stamped data points. The portable chlorine meters are compatible with Smart Link 2 software allowing the unit to interface with a PC and download data for analysis. The portable chlorine meters have an easy to use menu driven system displayed on an easy to read graphic LCD display.
Portable Chlorine Meters Calibration
The portable chlorine meters have three different measurement modes, chlorine, turbidity, and colorimetric testing. The portable chlorine meter includes 100 DPD 1 and 3 chlorine reagent tablets allowing you to perform 100 chlorine test measurements. Global Water offers liquid chlorine reagents in 60mL bottles, which will allow you to make an additional 250 chlorine measurements.
For turbidity measurements the portable chlorine meter should be calibrated using the included turbidity standards. The portable chlorine meter’s turbidity standards have a one-year shelf life, however you should calibrate your turbidity measurements every four to five days. The recommended method of calibration is to pour 10mL of standard into one of the four provided testing vials and label it appropriately. This 10mL sample can be used for approximately one month before needing to be replaced.
The portable chlorine meter’s colorimetric measurement mode has been calibrated with colored standards of known concentrations of platinum cobalt. One unit of color is equivalent to the color that is produced by 1 mg platinum/L in the form of the chloroplatinate ion. You will need to initialize the portable chlorine meter’s colorimetric test by using a colorless blank prior to inserting your test vial.
Portable Chlorine Meters Specifications
Turbidity:
Unit of Measure: NTU, FNU, FAU, ASBC, & EBC
Range: 0-4000
Resolution: 0.01 NTU/FNU (0.00-10.99)
0.1 NTU/FNU (110-4000)
1 NTU/FNU (110-4000)
Reproducibility: 0.02 NTU/FNU, 0.5 FAU
Chlorine:
Range: 0-10 ppm
Resolution: 0.01 ppm (0-5)
0.1 ppm (5-10)
Color:
Range: 0-500 CU (color units)
Resolution: 0.1 CU (0-100 CU)
1 CU (100-500 CU)
Light Source: 860nm LED (ISO)/Tungsten (EPA), 525 nm LED , UV LED 375
Accuracy: + 2%, 0.02 or + 2%, ±0.5 CU or 2%
Detection Limit: 0.05 NTU/FNU, 0.02 ppm, 0.2 CU
Signal Averaging: Disabled, 2, 5, 10
Power: 9V Battery, AC Optional
Data Logging: 4000 points
Auto Shut-off: Disabled, 5, 10, 30
Response time: <5 s
Languages: English, French, Spanish, Japanese, Italian, & Portuguese
Size: 8-7/8”x4-5/8”-1-7/16” (226mmx117mmx36mm)
Weight: 3 lbs (1250g)
|
Other Product
|